HUVY featured in Capital's special report on AI made in France

HUVY featured in Capital's special report on AI made in France

October 2026

Artificial Intelligence

Entrepreneurship

HUVY is among the solutions featured in the special report “The Unsung Geniuses of Made-in-France AI”, published by Capital magazine. Authored by Stéphane Barge, it highlights a new generation of French entrepreneurs leveraging artificial intelligence to address major everyday challenges: healthcare, the environment, education, and climate.

This selection shares a common ambition: making AI a lever for positive impact and committing to the long term. For HUVY, inclusion in this report is a recognition of the entire team’s dedication to improving access to dermatology.

We thank Capital for this feature.

However, certain statements in the article require clarification, in the best interest of both patients and healthcare professionals.


Clarification
HUVY is a CE-certified Class IIb medical device that assists healthcare professionals in evaluating melanoma risk. Capital magazine recently dedicated an article to us in its special report on artificial intelligence, based on an interview with Léonie Schröder, President and Co-founder. We thank the editorial staff at Capital and Stéphane Barge for their interest in our solution. Nonetheless, two statements in this article exceed our stated claims. Here is what our data substantiate.

The article states that "HUVY generates no false negatives and that, when presented with a green result, the patient can be certain of the benignity of their lesion." 
HUVY claims a sensitivity of 96% and a specificity of 72%. No device can guarantee the total absence of false negatives, and a green result is neither a diagnosis nor a certainty. Regardless of the result, any lesion that changes or causes concern must be examined by a healthcare professional.

Our Certified Performance
Under its Class IIb CE marking, HUVY demonstrates a sensitivity of 96% and a specificity of 72%, established on over 3,000 lesions and published in JEADV Clinical Practice.

In concrete terms:

  • 96% Sensitivity: Among at-risk lesions, 96% are correctly identified by HUVY. A false negative—meaning an at-risk lesion erroneously classified as low risk—remains possible. This is the case for any risk assessment support device.

  • 72% Specificity: Among benign lesions, 72% are correctly classified as low risk. The remainder may be referred to a dermatologist as a precaution. HUVY is designed to prioritize caution.

Our Post-Market Surveillance Studies
Two studies conducted since commercialization identified no false negatives among the analyzed cases. While this finding is encouraging, it applies specifically to these sample cohorts and does not constitute a guarantee of zero false negatives.

What a Green Result Means
HUVY displays its output using a color code. A green result (low risk) prompts a recommendation for monitoring. It constitutes neither a diagnosis nor a certainty: regardless of the result, any lesion that changes in appearance or causes concern must be examined by a healthcare professional. In addition, self-monitoring is explicitly encouraged for HUVY patients.

The Role of HUVY
HUVY does not make diagnoses. The device qualifies a risk level to support the healthcare professional's decision-making. The clinical decision remains with the practitioner, and the definitive diagnosis rests with the dermatologist.

We have submitted these clarifications to the editorial staff.